Tube Voltage: 60kV
Tube Current: 1.5mA
Exposure Time: 0.1--2.0s
Tube Focus: 0.3*0.3mm
Frequency: 30kHz
1. The body is small, lightweight, and has nearly no radiation.
2. It has excellent image quality, portable storage, saves more space.
3. It uses a high frequency and DC international power supply.
4. All of the components installed in the central PC board are concentrated. Shock, set-up, electron tubes, all of them are insulation vacuum, sealed stereotype protection.
5. There are also manual buttons installed on the surface of the shell, as well as batteries and charges.
6. This unit mainly suitable for oral pre-treatment for learning internal organizational structure, root depth, and so on, is indispensable in daily life clinic equipment, especially for dental implant surgery.
7. It also can connect with sensors, which is a great convenience.
8. The battery is durable, it can take about five hundred images after fully charged, and it can be charged and discharged for one thousand times in its life.
Tube Voltage | 60kV | Tube Current | 1.5mA |
Exposure Time | 0.1--2.0s | Tube Focus | 0.3*0.3mm |
Frequency | 30kHz | Battery | DC14.8V 6400mA |
Rated Power | 60KV | Charger Input | AC100V-240V±10% |
Charger Output | DC16.8V | Ambient temperature | 5℃~40℃ |
Distance from skin to cone | 130mm | Relative temperature | <75%RH |
1. In order to ensure safety in operation and use, good ground lead must be installed well.
2. Because small X-ray machine's thermal capacity is limited, attention should be paid to making the machine work intermittently.
3. Only during exposure process this machine will produce X-rays. It won't produce X-rays in power-off state, neither in power-on state as long as not in exposure.
4. If any unforeseen circumstances occur during correct use, operators should cut off the power immediately and stop using the machine. It cannot be used again until it is overhauled and qualified.
5. Before leaving the factory, each X-ray device is equipped with an additional filter plate with the following specifications:1.0/0.5mm aluminum sheets.
6. Every time when disassembling the machine, disassemble the head first and then other parts.
7. Every time after using the machine, recover the cross arm to the highest point, and then place the machine head in an equilibrium position.
8. Every time after using, sterilize and clean beam tube end that contacts the patient with medical alcohol.
9. Clean the machine surface with a dry cloth every week and inspect whether the connectors contact well.
10. Pay attention to machine maintenance. Check whether the indicator lights, display, rotary knob, switch, and remote control work normally; whether the X-ray source modules leak oil; whether when X-ray source modules expose produces any abnormal noise; or whether fixed screws in the mechanical part loose, and the like.
11. The rated capacity of the machine is 1.1 kw. Please select power supply or power grid with a power supply capacity greater than 1.5kw. If network voltage in this region is abnormal, we recommend matching 1.5 KW or higher industrial voltage stabilizer, to prevent damage to the machine and reduced machine life due to abnormal voltage.